| |
Links in this
section:
Introduction Fundamentals of the TEM technique Beam-sample interaction The Analytical TEM Detector Protection Qualitative Analysis Quantitative Analysis Microanalysis Examples (1) Microanalysis Examples (2) Microanalysis Examples (3) Summary
|
The Analytical TEM (AEM)
Analytical TEMs are commonly equipped with a variety of detectors for sample
analysis. The most important tools are bright field and dark field imaging,
electron diffraction, electron energy loss spectroscopy (EELS), and EDS. Due
to the confined space in the sample – polepiece region, EDS detector
positioning and collimation are of utmost importance. The electron column
must be designed to minimize stray radiation passing down the column and
generating X-rays from microscope components that may cause artefacts in the
EDS spectrum. Analytical TEMs are equipped with special apertures to
eliminate hard X-rays being generated in the column and polepiece.
The presence of stray X-rays can be determined by performing a “hole count
test”. This is performed by positioning the focussed electron beam in such a
way that it is transmitted through a hole in the specimen. The specimen
itself may scatter electrons and increase the stray radiation so the hole
count test doesn’t represent the true conditions during analysis. Therefore,
an even better test is to use a specially prepared standard sample (e.g.
NiOx) and check the level of spurious contributions to the spectrum.
Detector geometry for TEMs presents a special challenge for designers due to
the inherent space limitations in the objective polepiece area where the
sample resides (Figure 3). Ideally, the detector should be able to view
X-rays from above the specimen so that the sample can be analyzed in a
horizontal or near horizontal position. The position of the detector is
usually described by the take off angle (TOA) relative to the horizontal
specimen plane. Care should be taken to limit the entry of backscattered
electrons, as magnetic electron traps cannot be used in the TEM, due to the
fact that they would cause astigmatism in the objective lens. Consequently,
design of the collimator is very important. The solid angle of collection
can be improved by increasing detector area although this tends to worsen
the resolution of the detector. Alternatively the detector can be moved
closer to the specimen but this presents severe challenges for a design that
also has to minimize the pickup of stray radiation. TEM detectors are often
fitted with a 30mm2 area detector that gives reasonable
resolution performance and 0.1-0.3sr solid angle when positioned 10-20mm
from the specimen. This is significantly higher than the solid angle in a
typical SEM. The solid angle is most important in AEMs as the volume of the
material being analyzed is much smaller than that in a typical SEM analysis,
hence far less X-rays will be generated. This is especially important in the
analysis of biological materials. Peaks from the light elements B,C,O,N will
often be overlapped by the L and M peaks of heavier elements. This overlap
can be corrected by software if the peak shapes are well known. However,
incomplete charge collection (ICC) in the detector can distort low energy
peak shapes and it is therefore vital to determine what performance the
detector will achieve in this low energy range if light elements are to be
successfully detected. This is typically measured using the
ISO 15632:2002
recommendation of measuring resolution at FKα and CKα.
Back | Next
|