| |
Links in this
section:
Introduction Evolution of WDS technique Basics of WDS Diffraction Crystals Detectors & Geometry Comparison of EDS and WDS Qualitative Analysis Quantitative Analysis Mapping Summary
|
Qualitative Analysis
Using normal ED operating
conditions it is not always easy to positively identify peaks due to trace
elements. However, by increasing the beam current and slowly scanning over
the peak position using the WD spectrometer, trace elements can be reliably
identified.
Fig 6. shows the ED spectrum from
an alloy containing 0.15wt% Si. The red line shows the expected peak
position for Si, but it is difficult to be positive about reliable
identification. In the WD spectrum (fig 7), the improvement in peak to
background ratio means that there is no doubt that Si is present .
Back | Next |