| |
Navigation to
sections:
The Microanalysis System
What makes a Good Detector?
The Pulse Processor
Links in this
section
Role of the pulse processor Analog pulse shaping Time variant shaping Digital pulse shaping Fixed process time Adaptive pulse shaping Resolution & count rate Pulse pile-up protection Comparing different pulse processors Summary
|
The pulse
processor
Digital
pulse shaping
In a processor using digital
shaping, the signal from the preamplifier is digitized at the input of the
pulse processor, and shaping and noise reduction are achieved by digital
computation. The preamplifier output is sampled continuously by an analog to
digital converter (ADC) and X-ray pulse heights are measured by subtracting
the average of one set of values, measured before an X-ray event, from that
for another set, measured after the event. The resultant value of the step
measurement is then sent directly to the computer multichannel analyzer. The
noise on the voltage ramp from the detector is effectively filtered out by
averaging the signal (Fig. 15). The time over which the waveform is averaged
(often called the process time) is equivalent to the peaking time for an
analog shaper (TP).
By measuring the voltage ramp in
this way the benefits of perfect baseline recovery are achieved without the
complexities required for time variant shaping circuits. Therefore peaks
should not shift with count rate. There is also the potential for measuring
the zero baseline (‘strobing’) by averaging the digital output of the ramp
when no events are present (Fig. 18a). This will provide not only automatic
correction for changes in the slope of the voltage ramp caused by changes in
detector leakage current, but will also measure the effective noise
resolution. Calibration procedures for processors using digital pulse
shaping should be straightforward and reliable. One energy calibration
should be sufficient to guarantee accurate energy calibration at all count
rates.
Back | Next |