Navigation to sections:

 

The Microanalysis System

What makes a Good Detector?

The Pulse Processor

 

Links in this section

 

Role of the pulse processor
Analog pulse shaping
Time variant shaping
Digital pulse shaping
Fixed process time
Adaptive pulse shaping
Resolution & count rate
Pulse pile-up protection
Comparing different pulse processors
Summary

 

The pulse processor

 

Analog pulse shaping

 

In analog shaping the signal from the preamplifier is converted to a pulse by way of analog shaping electronics (Fig. 14). The height of the pulse is then measured and converted into a digital signal by an analog to digital converter (ADC). A longer peaking time (TP) is used to reduce noise and improve resolution.

 

Accurate height measurement of each pulse requires the shaping circuitry to be reset to the baseline level before another event can be accepted for measurement. The decay time (TD) gives the effective minimum time before another step can be measured accurately (Fig. 14).

 

 

The challenge with analog shaping is to compensate for changes in the baseline. To measure a 10keV photon to within 1eV requires the baseline to be known within 0.01% of the pulse height. The time required for the pulse to decay to below 0.01% of the pulse height will be many times the peaking time. As the count rate increases the residuals on the baseline left over from each shaped pulse begin to build up, resulting in a measurable shift in the apparent baseline. In addition, the baseline will move at high input rates due to the cumulative after-effects of all the large restore signals. Thus, as the input count rate increases, peaks will shift in position.

 

Even at low count rates, the slope on the voltage ramp will affect the position of the baseline. This slope, caused by leakage current from the detector (see Fig. 4), will mean that shaped pulses will not return exactly to zero, and therefore the baseline always has some offset.

 

When calibrating systems with analog shaping two key adjustments are required. One to compensate for detector leakage current, another to compensate for count rate-induced peak shift. Since the degree of shift depends on the shaping time, these adjustments have to be performed for every shaping time constant and gain setting. Count rate effects may also vary with spectrum content and even if exact compensation is achieved for an ‘average’ sample spectrum, peak shifts of a few eV may still be expected when different samples are analyzed. Furthermore, if detector leakage changes, the calibration process for baseline offset has to be repeated.

 

Some pulse processors are described as digital, because they use digital circuits and software instead of control knobs to control the circuitry. These processors will have the same problems with baseline measurement and calibration shown by any pulse processor that uses analog shaping.

 

Back | Next