| |
Navigation to
sections:
The Microanalysis System
What makes a Good Detector?
The Pulse Processor
Links in this
section:
Manganese resolution Low energy resolution Fluorine resolution Incomplete charge collection Carbon resolution Fe55 source Changes with time Geometry Summary
|
What
makes a good detector?
Detector
specifications based on tests using an Fe55 radioactive source
Testing the performance of an EDS
detector using a radioactive source is convenient for a detector
manufacturer because it can be done without having to mount a detector on an
electron microscope column. However, Fe55 source specifications don’t
necessarily guarantee the performance when mounted on a column and
collecting X-rays emitted during electron bombardment. They are also limited
in scope and do not reveal some important aspects of detector performance,
in the real situation of electron microscope based X-ray microanalysis.
The
reason that the resolution of a detector is traditionally specified for
Manganese Kα X-rays at 5.895 keV, is because this is the energy of the most
intense X-ray line emitted by the Fe55 source (Fig. 10).
One useful feature
of the Fe55 source is the absence of the continuum or bremsstrahlung X-rays
that would be generated in an electron microscope. Therefore, in an Fe55
spectrum the very low intensity background at lower energies than the
characteristic Mn X-ray peaks is due to incomplete charge collection. The
most common measurement of ICC is the peak: background ratio comparing the
height of the MnKα peak to the average background between 0.9-1.1keV (Fig.
10). Events appearing at 1keV correspond to MnKα photons where 83% of the
charge has not been collected, whereas events appearing at say 4keV have
lost 32% of the charge. Although the original IEEE standard suggested
measurement at several energies, manufacturers have most commonly used 1keV.
As shown in Fig. 10, 1keV is typically the lowest part of the ICC background
and a low value at this energy (high peak:background ratio) does not
guarantee that charge collection will be good for the smaller charge losses
that contribute to the tail or plateau on the low energy side of a peak.
Fe55 peak:background only gives a general indication of charge collection
efficiency, but does not ensure good peak shapes or give any indication of
ICC for low energy X-rays. Incomplete charge collection is most important
where X-rays are strongly absorbed and only penetrate a short distance into
the crystal, for example just above the absorption edge of the elements
making up the crystal (1.84keV for Si(Li) detectors), and for very low
energy X-rays (less than 0.5keV). The CKα resolution measurement on the
microscope represents a more useful and sensitive measurement of incomplete
charge collection than peak:background ratios measured with an Fe55 source.
Low values for CKα FWHM guarantee excellent charge collection, and good peak
shapes at all energies.
Back | Next |