| |
Navigation to
sections:
The Microanalysis System
What makes a Good Detector?
The Pulse Processor
Links in this
section:
Manganese resolution Low energy resolution Fluorine resolution Incomplete charge collection Carbon resolution Fe55 source Changes with time Geometry Summary
|
What
makes a good detector?
Incomplete
charge collection
If all the electron-hole pairs
generated by an X-ray are not swept to the electrical contacts, the charge
signal measured by the FET will be lower than expected, and the energy
measured lower than the energy of the incident X-ray. This phenomenon is
known as incomplete charge collection (ICC), and results in counts appearing
in the spectrum at lower energies than the energy of the X-ray which they
represent, typically as a tail on the low energy side of the peak. All
detectors suffer from incomplete charge collection to some extent. Low
energy X-rays have a very shallow depth of penetration and ICC is usually
poor near the front contact. The peak for a low energy X-ray will therefore
be broader and have a mean energy lower than expected, due to varying levels
of charge collection as each X-ray is measured
(Fig. 9). Thus, incomplete
charge collection results in detectors with resolutions measured at low
energy that are worse than those predicted by theory. In extreme cases it
can be the dominant factor controlling resolution at these very low energies
(Fig. 9).
Back | Next |