Navigation to sections:

 

The Microanalysis System

What makes a Good Detector?

The Pulse Processor

 

Links in this section:

 

Manganese resolution
Low energy resolution
Fluorine resolution
Incomplete charge collection
Carbon resolution
Fe55 source
Changes with time
Geometry
Summary

 

What makes a good detector?

 

Fluorine resolution FKα FWHM

 

The width of the peaks in a spectrum will vary depending on the energy of the X-ray line. In Fig. 8 this variation with energy has been calculated to show how the resolution should change with X-ray energy for Si(Li) and HpGe detectors with different Mn resolution specifications. The curves are calculated from the equation

 

FWHM2 = k . E + FWHMnoise2

 

where k is a constant for the detector material, and E is the energy.

 

These curves demonstrate that as the energy decreases, the resolution of the X-ray peaks improves. The variation is also clearly different for Si(Li) and HpGe crystals. At low energy the electronic noise contribution (FWHMnoise) has a greater effect on resolution (Fig. 9). Mn FWHM is a very insensitive measure to characterize the noise of a detector and predict the resolution at low energy. Therefore to characterize the resolution of a detector at low energy, the resolution is also quoted for another line, typically fluorine Kα.

 

Back | Next