| |
Navigation to
sections:
The Microanalysis System
What makes a Good Detector?
The Pulse Processor
Links in this
section:
Manganese resolution Low energy resolution Fluorine resolution Incomplete charge collection Carbon resolution Fe55 source Changes with time Geometry Summary
|
What
makes a good detector?
The
importance of low energy resolution
The identification and
quantification of closely spaced X-ray peaks becomes easier and more
accurate as the separation between them increases. X-ray lines get closer
together at low energy and this is apparent in Fig. 6 where the energy of
the main alpha line for K, L and M series is plotted for all elements. At
the MnKα energy most commonly used to specify EDS resolution, peaks are well
separated. However, it is clear that much more serious overlaps occur below
3keV and the resolution performance at low energies is critical to good
performance for all elements.
When small features <1µm in size are being analyzed, the beam voltage needs
to be reduced to avoid electron scattering outside the feature. However, at
low kV only low energy lines are available for analysis. Spectra collected
from a nickel alloy at 20kV and 5kV (Fig. 7) illustrate the importance of
resolution at low energy. When working at 20kV, the separation of widely
spaced K lines of Cr, Fe and Ni, will not be affected much by a few eV
variation in resolution. When working at 5kV however, where identification
relies on L lines, a detector with a few eV better resolution will allow the
L lines of Cr, Fe and Ni to be separated enough for confident
identification.
Back | Next |